1. Four pi calibration and modeling of a bare germanium detector in a cylindrical field source.
- Author
-
Dewberry, R. A. and Young, J. E.
- Subjects
- *
GERMANIUM , *DETECTORS , *MATHEMATICAL models , *MONTE Carlo method , *PARTICLES (Nuclear physics) - Abstract
In this paper we describe a 4π cylindrical field acquisition configuration surrounding a bare (unshielded, uncollimated) high purity germanium detector. We perform an efficiency calibration with a flexible planar source and model the configuration in the 4π cylindrical field. We then use exact calculus to model the flux on the cylindrical sides and end faces of the detector. We demonstrate that the model accurately represents the experimental detection efficiency compared to that of a point source and to Monte Carlo N-particle (MCNP) calculations of the flux. The model sums over the entire source surface area and the entire detector surface area including both faces and the detector's cylindrical sides. Agreement between the model and both experiment and the MCNP calculation is within 8%. [ABSTRACT FROM AUTHOR]
- Published
- 2012
- Full Text