1. Development of a high energy resolution and wide dose rate range portable gamma-ray spectrometer.
- Author
-
Wang Y, Feng C, Zhao M, Shan C, Liu F, Lei Q, Zhou Z, and Liu S
- Subjects
- Gamma Rays, Tellurium, Radiometry, Neutrons
- Abstract
In this paper, a portable gamma-ray spectrometer for real-time and in-situ gamma-ray detection applications is presented. By combining a quasi-hemispherical CdZnTe (CZT) semiconductor detector and a Geiger-Muller (GM) counter together, a wide dose rate range is achieved, ranging from 0.1 μSv/h to 100 mSv/h with a relative error of less than 10%. The GM counter is used to measure dose rate from 1 mSv/h to 100 mSv/h. With CZT, the spectrometer can provide a high energy resolution spectrum for nuclide identification and a high precision dose rate at low dose rates. The full width half maximum (FWHM) resolution is 2.2% at 662 keV below 70 μSv/h and is better than 3.3% at 3.8 mSv/h. The weight of the spectrometer is 3.2 kg for handheld and the runtime is up to 12 h without charging. For preliminary applications, the spectrometer was used to measure the gamma radiation around the Back-n white neutron beam line at China Spallation Neutron Source and around the steam generator in the nuclear power plant at Daya Bay Nuclear Power Station., Competing Interests: Declaration of competing interest The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper., (Copyright © 2022 Elsevier Ltd. All rights reserved.)
- Published
- 2023
- Full Text
- View/download PDF