1. Clinical performance comparison of two medium cut-off dialyzers.
- Author
-
Krieter DH, Rüth M, Lemke HD, and Wanner C
- Subjects
- Humans, Myoglobin, Albumins, Cephalosporins, Renal Dialysis instrumentation
- Abstract
Introduction: Medium-cut-off (MCO) dialyzers may beneficially impact outcomes in patients on hemodialysis., Methods: In a randomized, controlled trial in maintenance hemodialysis patients, the new Nipro ELISIO-17HX MCO dialyzer was compared to the Baxter Theranova 400 filter regarding middle molecule removal. Furthermore, the suitability of two assays for free lambda-light chain (λFLC) detection (Freelite vs. N-Latex) was verified., Results: ELISIO-HX achieved slightly lower reduction ratios for β
2 -microglobulin (71.8 ± 6.0 vs. 75.3 ± 5.8%; p = 0.001), myoglobin (54.7 ± 8.6 vs. 64.9 ± 8.7%; p < 0.001), and kappa-FLC (62.1 ± 8.8 vs. 56.3 ± 7.7%; p = 0.021). λFLC reduction ratios were more conclusive with the Freelite assay and not different between ELISIO-HX and Theranova (28.4 ± 3.9 vs. 38.7 ± 13.4%; p = 0.069). The albumin loss of Theranova was considerably higher (2.14 ± 0.45 vs. 0.77 ± 0.25 g; p = 0.001) and the Global Removal ScoreLoss alb largely inferior (30.6 ± 7.4 vs. 82.4 ± 29.2%/g; p = 0.006) to ELISIO-HX., Conclusions: The new ELISIO-HX expands the choice of dialyzers for MCO hemodialysis., (© 2022 The Authors. Therapeutic Apheresis and Dialysis published by John Wiley & Sons Australia, Ltd on behalf of International Society for Apheresis and Japanese Society for Apheresis.)- Published
- 2023
- Full Text
- View/download PDF