medicine.medical_specialty ,Materials science ,X ray diagnosis ,business.industry ,Image quality ,Radiography ,Radiation dose ,Resolution (electron density) ,Imaging phantom ,Surgery ,Imaging quality ,medicine ,Radiology, Nuclear Medicine and imaging ,business ,X-Ray Intensifying Screens ,Biomedical engineering - Abstract
PURPOSE: This paper compares a new film-screen system (FSS) called INSIGHT Skeletal Imaging System with the previously used Lanex/T-MAT G FSS. MATERIAL AND METHODS: Using a Bronder phantom, measurements were made of dose, resolution and contrast. 135 skeletal phantom images were assessed in order of quality by six observers. RESULTS: Comparable high resolution film-screen combinations (FSC) showed similar geometric resolution. Comparing high intensifying screens, the new INSIGHT Skeletal Regular FSS showed better resolution than the Lanex medium FSC. Dose reduction for the INSIGHT Skeletal Imaging FSS was 29-56%. The new FSS showed image quality similar to high resolution screens but was significantly better when using high intensifying screens. CONCLUSION: The new INSIGHT Skeletal Imaging System can replace the Lanex/T-MAT G FFS by retaining quality but reducing radiation dose by 29-56%. Using the new high intensifying FSC, images showed in addition improved film quality.