1. Validation of the doubly labeled water method using off-axis integrated cavity output spectroscopy and isotope ratio mass spectrometry.
- Author
-
Melanson EL, Swibas T, Kohrt WM, Catenacci VA, Creasy SA, Plasqui G, Wouters L, Speakman JR, and Berman ESF
- Subjects
- Adult, Calorimetry, Indirect methods, Deuterium chemistry, Deuterium urine, Energy Metabolism, Exercise Test, Female, Humans, Male, Middle Aged, Oxygen Isotopes urine, Spectrum Analysis methods, Young Adult, Isotope Labeling methods, Mass Spectrometry methods, Oxygen Isotopes chemistry, Pulmonary Gas Exchange physiology, Water chemistry
- Abstract
When the doubly labeled water (DLW) method is used to measure total daily energy expenditure (TDEE), isotope measurements are typically performed using isotope ratio mass spectrometry (IRMS). New technologies, such as off-axis integrated cavity output spectroscopy (OA-ICOS) provide comparable isotopic measurements of standard waters and human urine samples, but the accuracy of carbon dioxide production (V̇co
2 ) determined with OA-ICOS has not been demonstrated. We compared simultaneous measurement V̇co2 obtained using whole-room indirect calorimetry (IC) with DLW-based measurements from IRMS and OA-ICOS. Seventeen subjects (10 female; 22 to 63 yr) were studied for 7 consecutive days in the IC. Subjects consumed a dose of 0.25 g H2 18 O (98% APE) and 0.14 g2 H2 O (99.8% APE) per kilogram of total body water, and urine samples were obtained on days 1 and 8 to measure average daily V̇co2 using OA-ICOS and IRMS. V̇co2 was calculated using both the plateau and intercept methods. There were no differences in V̇co2 measured by OA-ICOS or IRMS compared with IC when the plateau method was used. When the intercept method was used, V̇co2 using OA-ICOS did not differ from IC, but V̇co2 measured using IRMS was significantly lower than IC. Accuracy (~1-5%), precision (~8%), intraclass correlation coefficients ( R = 0.87-90), and root mean squared error (30-40 liters/day) of V̇co2 measured by OA-ICOS and IRMS were similar. Both OA-ICOS and IRMS produced measurements of V̇co2 with comparable accuracy and precision compared with IC.- Published
- 2018
- Full Text
- View/download PDF