1. Quantification of plasma homocitrulline using hydrophilic interaction liquid chromatography (HILIC) coupled to tandem mass spectrometry.
- Author
-
Jaisson, Stéphane, Gorisse, Laëtitia, Pietrement, Christine, and Gillery, Philippe
- Subjects
- *
AMINO acids , *LYSINE , *BIOMARKERS , *CHRONIC kidney failure , *ATHEROSCLEROSIS , *HYDROPHILIC interaction liquid chromatography - Abstract
Homocitrulline (HCit), an amino acid formed by the carbamylation of ε-amino groups of lysine residues, is considered a promising biomarker for monitoring diseases such as chronic renal failure and atherosclerosis. This paper describes a tandem mass spectrometric method for total, protein-bound and free HCit measurement in plasma samples. HCit was separated from other plasma components by hydrophilic interaction liquid chromatography. Detection was achieved by monitoring transitions of 190.1 > 127.1 and 190.1 > 173.1 for HCit, and 183.1 > 120.2 for d-citrulline used as internal standard. This method allowed HCit quantification within 5.2 min and was precise (inter-assay CV < 5.85%), accurate (mean recoveries ranging from 97% to 106%), and exhibited a good linearity from 10 nmol/L to 1.6 μmol/L. Plasma samples from control and uremic mice ( n = 10) were analyzed. In control mice, mean total plasma HCit concentration was 0.78 ± 0.12 μmol/mol amino acids, whereas it was increased 2.7-fold in uremic mice plasma, reaching 2.10 ± 0.50 μmol/mol amino acids ( p < 0.001). In conclusion, this method exhibits good analytical performances and meets the criteria of sensitivity suitable for HCit concentration assessment in plasma samples. [ABSTRACT FROM AUTHOR]