1. A new immunohistochemical ER/PR image analysis system: a multisite performance study.
- Author
-
Nassar A, Cohen C, Agersborg SS, Zhou W, Lynch KA, Heyman ER, Olson A, Lange H, and Siddiqui MT
- Subjects
- Double-Blind Method, Feasibility Studies, Female, Humans, Immunohistochemistry methods, Mammary Glands, Human pathology, Microscopy, Observer Variation, Practice Guidelines as Topic, Receptors, Estrogen immunology, Receptors, Estrogen metabolism, Receptors, Progesterone immunology, Receptors, Progesterone metabolism, Image Processing, Computer-Assisted, Mammary Glands, Human metabolism
- Abstract
Background: Aperio provides a new image analysis (IA) solution for immunohistochemistry (IHC) as part of its digital pathology system. To be used in a clinical setting, substantial equivalence to scoring by manual microscopy (MM) needs to be shown. A multisite study was conducted to assess the performance of Aperio's IHC IA solution for estrogen receptor (ER) and progesterone receptor (PR)., Design: A total of 260 formalin-fixed, paraffin-embedded breast tissue specimens were assayed at 2 clinical sites for ER and PR. The ability to score ER/PR slides in terms of (1) percentage of positive nuclei with cutoffs of 1%, 5%, and 10% and (2) average staining intensity as 0, 1+, 2+, and 3+ score was assessed. At each site, 3 pathologists performed a blinded read of the glass slides using their microscopes. The glass slides were then scanned, and after a wash-out period and randomization of the slides, the pathologists viewed the images on a computer monitor and outlined a representative set of tumor regions to be analyzed by IA. Each of the methods: MM and IA were evaluated separately and comparatively., Results: Comparable or higher percent agreements were obtained for IA compared with MM (ER--percent of positive nuclei with cutoffs: MM: 91.3% to 98.8%/IA: 93.8% to 98.8%/IA vs. MM: 92.5% to 97.5%, and intensity score: MM: 55.0% to 86.3%/IA: 88.8% to 90.0%/IA vs. MM: 63.8% to 86.3%; PR-percent of positive nuclei with cutoffs: MM: 83.8% to 99.0%/IA: 85.0% to 99.0%/IA vs. MM: 81.3% to 99.0%, and intensity score: MM: 58.8% to 88.0%/IA: 68.8% to 88.0%/IA vs. MM: 58.8% to 84.0%)., Conclusions: The study results show that Aperio's digital IHC IA solution for ER/PR is substantially equivalent to scoring by MM.
- Published
- 2011
- Full Text
- View/download PDF