1. Combined anterior segment OCT and wavefront-based autorefractor using a shared beam
- Author
-
Yu Cherng Chang, Heather Ann Durkee, Florence Cabot, Ettore Masetti, Sonia H. Yoo, Marco Ruggeri, Jean-Marie A. Parel, Fabrice Manns, Giulia Belloni, and Arthur Ho
- Subjects
Wavefront ,Refractive error ,Accuracy and precision ,Ophthalmic surgical microscope ,medicine.diagnostic_test ,genetic structures ,Image quality ,business.industry ,Computer science ,medicine.disease ,Atomic and Molecular Physics, and Optics ,eye diseases ,Article ,Optics ,Optical coherence tomography ,Autorefractor ,medicine ,sense organs ,business ,Beam (structure) ,Biotechnology - Abstract
We have combined an anterior segment (AS) optical coherence tomography (OCT) system and a wavefront-based aberrometer with an approach that senses ocular wavefront aberrations using the OCT beam. Temporal interlacing of the OCT and aberrometer channels allows for OCT images and refractive error measurements to be acquired continuously and in real-time. The system measures refractive error with accuracy and precision comparable to that of clinical autorefractors. The proposed approach provides a compact modular design that is suitable for integrating OCT and wavefront-based autorefraction within the optical head of the ophthalmic surgical microscope for guiding cataract surgery or table-top devices for simultaneous autorefraction and ocular biometry.
- Published
- 2021