1. Taking advantage of a high-throughput flow cytometer for the implementation of an ADCC assay for regulatory compliance
- Author
-
Rosa Camacho-Sandoval, Alexis Jiménez-Uribe, Alejandra V. Tenorio-Calvo, Carlos A. López-Morales, Leslie Muñoz-García, Alejandra Montes-Luna, Héctor Leonardo García-Xolalpa, Marco Velasco-Velázquez, Lenin Pavón, Sonia Mayra Pérez-Tapia, and Emilio Medina-Rivero
- Subjects
ADCC ,High-throughput flow cytometry ,Validation bioassay ,Biotechnology ,TP248.13-248.65 - Abstract
Technological advances allowed the development of high-throughput instruments such as IntelliCyt iQue Screener PLUS®. Here, we took advantage of this technology to transfer a previously validated cytotoxicity assay. The evaluated parameters were cell permeability, caspase activation and phosphatidyl serine exposure. The assay was accurate (r2 = 0.90), precise (%CV ≤ 18.90) and specific. These results showed that this technology is suitable to be used in control quality environments. In addition, the automation provided a faster acquisition and analysis of data with precise and accurate results. This application could be implemented to evaluate another in vitro mechanism of action of different biotherapeutics.
- Published
- 2020
- Full Text
- View/download PDF