1. Respiration Monitoring System of Lung Phantom Using Magnetic Sensor
- Author
-
Rocky Alfanz, David Abraham, Imamul Muttakin, and Rohmadi Rohmadi
- Subjects
Telemonitoring ,Respiration monitoring ,Control and Optimization ,Computer Networks and Communications ,Computer science ,Conductive materials ,Respiratory pattern ,Continuous mode ,Imaging phantom ,Inductance ,Hardware and Architecture ,Control and Systems Engineering ,Lung phantom ,Computer Science (miscellaneous) ,Magnetic sensor ,Expiration ,Electrical and Electronic Engineering ,LDC1000 ,Instrumentation ,Simulation ,Information Systems - Abstract
Monitoring vital signs is substantial in healthcare to assist both diagnosis and treatment. This work proposes a means of telemonitoring system at initial stage to observe respiratory pattern on lung phantom. Magnetic sensor module LDC1000 is used to read inductance value of conductive material in relation to distance variation. Therefore, respiration pattern can be observed. In continuous mode, the inspiration inductance value is 8 uH, while expiration is 17 uH, with stoppage is 17 uH. For static measurement, the inspiration inductance value is 7.80 uH, while expiration is 16.46 uH and stoppage is 16.46 uH. Those values could be further referred for vital signs telemonitoring system design based on contactless and portable devices.
- Published
- 2018
- Full Text
- View/download PDF