1. How Controlled is the Expansion of VIATORR CX?
- Author
-
Dell T, Menne M, Wagenpfeil J, Praktiknjo M, Jansen C, Isaak A, Mesropyan N, Steinseifer U, Attenberger U, Luetkens J, Meyer C, and Kuetting D
- Subjects
- Humans, Stents, Prosthesis Design, Portasystemic Shunt, Transjugular Intrahepatic
- Abstract
Purpose: To investigate and compare the physical properties of the new generation Gore VIATORR-Controlled Expansion Endoprosthesis (VCX) to those of the predecessor VIATORR stent in an in vitro experimental setup., Materials and Methods: A total of 12 stents (8 VCX; 4 VIATORR; GORE, USA) were examined. Radial resistive force (RRF) and chronic outward force (COF) were assessed using a radial force testing machine (RX-650, Machine Solutions Inc., USA). To assess the radial forces of the VCX above 8 mm, balloon expansion was performed between cycles., Results: All VCX stents show an abrupt decrease in COF at an external diameter of 8.3 mm; RRF decreases likewise at an external diameter of 8.5 mm. The predecessor VIATORR stent without the "controlled expansion" feature shows linear radial force reduction until full expansion at a diameter of 10 mm. The physical properties of the VCX can be altered by balloon modulation. Point of COF (RRF) reduction shifts to 8.5 mm (8.6 mm), 8.6 mm (8.8 mm) and 9.3 mm (9.6 mm) following modulation with a 8 mm, 9 mm and 10 mm balloon., Conclusions: The VCX shows an abrupt and disproportionate decrease in COF and RRF at an external diameter of 8.3 mm, thus passive expansion to its nominal diameter of 10 mm is not to be expected. By means of balloon dilatation the physical properties of the stent can be altered, enabling customized TIPS creation. The previous VIATORR stent shows continuous COF and RRF until total expansion., (© 2023. The Author(s).)
- Published
- 2023
- Full Text
- View/download PDF