1. Pull-off resistance of a screwless implant-abutment connection and surface evaluation after cyclic loading.
- Author
-
Alevizakos V, Mosch R, Mitov G, Othman A, and von See C
- Abstract
Purpose: The aim of this study was to investigate to what extent cyclic load affects the screwless implant-abutment connection for Morse taper dental implants., Materials and Methods: 16 implants (SICvantage max) and 16 abutments (Swiss Cross) were used. The screwless implant-abutment connection was subjected to 10,000 cycles of axial loading with a maximum force of 120 N. For the pull-off testing, before and after the same cyclic loading, the required force for disconnecting the remaining 6 implant-abutment connections was measured. The surface of 10 abutments was examined using a scanning electron microscope 120× before and after loading., Results: The pull-off test showed a significant decrease in the vertical force required to pull the abutment from the implant with mean 229.39 N ± 18.23 before loading, and 204.30 N ± 13.51 after loading ( P <.01). Apart from the appearance of polished surface areas and slight signs of wear, no visible damages were found on the abutments., Conclusion: The deformation on the polished abutment surface might represent the result of micro movements within the implant-abutment connection during loading. Although there was a decrease of the pull-off force values after cyclic loading, this might not have a notable effect on the clinical performance., (© 2021 The Korean Academy of Prosthodontics.)