1. Impact of advanced detector technology and iterative reconstruction on low-dose quantitative assessment of lung computed tomography density in a biological lung model
- Author
-
Eric A. Hoffman, John D. Newell, Nicholas Stoyles, Kung-Sik Chan, Chelsea M. Sloan, Emily Hammond, Jessica C. Sieren, Junfeng Guo, and J. C. Ames
- Subjects
Percentile ,medicine.diagnostic_test ,Radon transform ,business.industry ,Image quality ,Detector ,General Medicine ,Iterative reconstruction ,computer.software_genre ,Air trapping ,Article ,030218 nuclear medicine & medical imaging ,03 medical and health sciences ,0302 clinical medicine ,030228 respiratory system ,Voxel ,medicine ,medicine.symptom ,Quantitative computed tomography ,Nuclear medicine ,business ,computer - Abstract
Purpose Quantitative computed tomography (QCT)-derived measures of lung density are valued methods for objectively characterizing lung parenchymal and peripheral airways disease and are being used in a growing number of lung disease focused trials. Detector and reconstruction improvements in CT technology have allowed for significant radiation dose reduction in image acquisition with comparable qualitative image quality. We report the impact of detector type and reconstruction type on QCT lung density measures in relation to decreasing dose indices. Methods Two sets of studies were completed in an in vivo pig model with a SOMATOM Definition Flash CT system: (a) prior to system upgrade with conventional detectors (UFC) and filtered back projection (FBP), and (b) post system upgrade with integrated electronic detectors (STELLAR) and iterative reconstruction (SAFIRE). CT data were acquired across estimated CT volume dose indices (CTDIvol ) ranging from 0.75 to 15 mGy at both inspiratory and expiratory breath holds. Semiautomated lung segmentations allowed calculation of histogram median, kurtosis, and 15th percentile. Percentage of voxels below -910 HU and -950 HU (inspiratory), and -856 HU (expiratory) were also examined. The changes in these QCT metrics from dose reduction (15 mGy down to 0.75 mGy) were calculated relative to paired reference values (15 mGy). Results were compared based on detector and reconstruction type. Results In this study, STELLAR detectors improved concordance with 15 mGy values down to 3 mGy for inspiratory scans and 6 mGy for expiratory scans. The addition of SAFIRE reconstruction in all acquired measurements resulted in minimal deviation from reference values at 0.75 mGy. Conclusion The use of STELLAR integrated electronic detectors and SAFIRE iterative reconstruction may allow for comparable lung density measures with CT dose indices down to 0.75 mGy.
- Published
- 2018