1. Catheter strategy to ease the procedure and reduce radiation exposure when requiring neck access
- Author
-
Brigitte Stiller, Rouven Kubicki, Johanna Hummel, René Höhn, Kevin Müller, and Jochen Grohmann
- Subjects
Diseases of the circulatory (Cardiovascular) system ,RC666-701 - Abstract
Objectives To assess the potential occupational radiation reduction and technical feasibility in patients rotated 180° (upside-down) when requiring neck access for transcervical or trans-subclavian catheterisation.Methods Upside-down positioning is defined as rotating patients in supine position by 180°, so that the feet come to rest where the head would otherwise be. We retrospectively evaluated all these procedures performed between March 2016 and May 2019. Furthermore, two different phantoms (paediatric and adult) were used prospectively to quantify the occupational dose between conventional or upside-down positioning. In this context, ambient dose equivalents were measured using real-time dosimeters. Three different projection angles were applied.Results 44 patients with median age and body weight of 1.0 year (range 0–56) and 9.5 kg (range 1.3–74.3) underwent 63 procedures positioned upside-down. This position proved advantageous for practical reasons, since the length of the examination table could be optimally used. Additionally, it resulted in a significantly lower overall ambient dose equivalent for the primary operator (PO) of 94.8% (mean: 2569±807 vs 135±23 nSv; p
- Published
- 2020