1. Doppler colour flow imaging and flow quantification with a novel forward-viewing intravascular ultrasound system
- Author
-
Gatzoulis, Loukianos, Ramnarine, Kumar V., Pye, Stephen D., Anderson, Tom, Newby, David E., Hoskins, Peter R., and McDicken, W.Norman
- Subjects
- *
INTRAVASCULAR ultrasonography , *CATHETERS - Abstract
The aim of this work was to investigate the potential of a novel forward-viewing intravascular ultrasound (IVUS) system for flow quantification and colour flow imaging combined with B-mode imaging. A stiff 3.8-mm diameter catheter was used to scan a 72° sector ahead of its tip. Operating at 30 MHz, the catheter was integrated with an IVUS scanner and a radiofrequency (RF) data-acquisition system. RF data were software processed for producing B-mode images and deriving velocity estimates. Steady flow in the range of 45 to 146 mL/min toward the catheter, was used in wall-less tissue-mimicking phantoms simulating healthy lumen (8-mm diameter), 30% diameter symmetrical stenosis and 37% diameter eccentric stenosis. The system provided colour flow images and good estimation of peak velocity and volumetric flows (within 1% to 9% and 16% to 48%, respectively, of calculated values) at 5 to 7 mm distal to the catheter. A sector forward-viewing IVUS imaging/Doppler system is suitable for combined anatomical and functional assessment of stenosed vessels. (E-mail: N.McDicken@ed.ac.uk) [Copyright &y& Elsevier]
- Published
- 2003
- Full Text
- View/download PDF