1. Automating quality control in cardiac magnetic resonance: Artificial intelligence for discriminative assessment of planning and motion artifacts and real-time reacquisition guidance
- Author
-
Hoi C. Cheung, Kavitha Vimalesvaran, Sameer Zaman, Michalis Michaelides, Matthew J. Shun-Shin, Darrel P. Francis, Graham D. Cole, and James P. Howard
- Subjects
Artificial intelligence ,Machine learning ,Cardiac magnetic resonance ,Quality control ,Quality assessment ,Convolutional neural networks ,Diseases of the circulatory (Cardiovascular) system ,RC666-701 - Abstract
Background: Accurate measurements from cardiovascular magnetic resonance (CMR) images require precise positioning of scan planes and elimination of motion artifacts from arrhythmia or breathing. Unidentified or incorrectly managed artifacts degrade image quality, invalidate clinical measurements, and decrease diagnostic confidence. Currently, radiographers must manually inspect each acquired image to confirm diagnostic quality and decide whether reacquisition or a change in sequences is warranted. We aimed to develop artificial intelligence (AI) to provide continuous quality scores across different quality domains, and from these, determine whether cines are clinically adequate, require replanning, or warrant a change in protocol. Methods: A three-dimensional convolutional neural network was trained to predict cine quality graded on a continuous scale by a level 3 CMR expert, focusing separately on planning and motion artifacts. It incorporated four distinct output heads for the assessment of image quality in terms of (a, b, c) 2-, 3- and 4-chamber misplanning, and (d) long- and short-axis arrhythmia/breathing artifact. Backpropagation was selectively performed across these heads based on the labels present for each cine. Each image in the testing set was reported by four level 3 CMR experts, providing a consensus on clinical adequacy. The AI's assessment of image quality and ability to identify images requiring replanning or sequence changes were evaluated with Spearman’s rho and the area under receiver operating characteristic curve (AUROC), respectively. Results: A total of 1940 cines across 1387 studies were included. On the test set of 383 cines, AI-judged image quality correlated strongly with expert judgment, with Spearman’s rho of 0.84, 0.84, 0.81, and 0.81 for 2-, 3- and 4-chamber planning quality and the extent of arrhythmia or breathing artifacts, respectively. The AI also showed high efficacy in flagging clinically inadequate cines (AUROC 0.88, 0.93, and 0.93 for identifying misplanning of 2-, 3- and 4-chamber cines, and 0.90 for identifying movement artifacts). Conclusion: AI can assess distinct domains of CMR cine quality and provide continuous quality scores that correlate closely with a consensus of experts. These ratings could be used to identify cases where reacquisition is warranted and guide corrective actions to optimize image quality, including replanning, prospective gating, or real-time imaging.
- Published
- 2024
- Full Text
- View/download PDF