1. Reproducibility and calibration of MMC-based high-resolution gamma detectors.
- Author
-
Bates, C. R., Pies, C., Kempf, S., Hengstler, D., Fleischmann, A., Gastaldo, L., Enss, C., and Friedrich, S.
- Subjects
- *
DETECTORS , *CALORIMETERS , *POLYNOMIALS , *CENTROID , *ANISOTROPY - Abstract
We describe a prototype γ-ray detector based on a metallic magnetic calorimeter with an energy resolution of 46eV at 60 keV and a reproducible response function that follows a simple second-order polynomial. The simple detector calibration allows adding high-resolution spectra from different pixels and different cool-downs without loss in energy resolution to determine γ-ray centroids with high accuracy. As an example of an application in nuclear safeguards enabled by such a γ-ray detector, we discuss the non-destructive assay of 242Pu in a mixed-isotope Pu sample. [ABSTRACT FROM AUTHOR]