1. Minimization of the Wilson's Central Terminal voltage potential via a genetic algorithm.
- Author
-
Moeinzadeh H, Bifulco P, Cesarelli M, McEwan AL, O'Loughlin A, Shugman IM, Tapson JC, Thiagalingam A, and Gargiulo GD
- Subjects
- Electrocardiography instrumentation, Humans, Algorithms, Electrocardiography methods, Electrophysiological Phenomena physiology, Signal Processing, Computer-Assisted
- Abstract
Objective: The Wilson Central Terminal (WCT) is an artificially constructed reference for surface electrocardiography, which is assumed to be near zero and steady during the cardiac cycle; namely it is the simple average of the three recorded limbs (right arm, left arm and left leg) composing the Einthoven triangle and considered to be electrically equidistant from the electrical center of the heart. This assumption has been challenged and disproved in 1954 with an experiment designed just to measure and minimize WCT. Minimization was attempted varying in real time the weight resistors connected to the limbs. Unfortunately, the experiment required a very cumbersome setup and showed that WCT amplitude could not be universally minimized, in other words, the weight resistors change for each person. Taking advantage of modern computation techniques as well as of a special ECG device that aside of the standard 12-lead Electrocardiogram (ECG) can measure WCT components, we propose a software minimization (genetic algorithm) method using data recorded from 72 volunteers., Result: We show that while the WCT presents average amplitude relative to lead II of 58.85% (standard deviation of 30.84%), our minimization method yields an amplitude as small as 7.45% of lead II (standard deviation of 9.04%).
- Published
- 2018
- Full Text
- View/download PDF