1. Transcatheter leadless pacemaker implantation in a patient with a transvenous dual-chamber pacemaker already in place.
- Author
-
Karjalainen PP, Nammas W, and Paana T
- Subjects
- Adult, Atrial Fibrillation complications, Atrioventricular Block complications, Cardiac Catheterization instrumentation, Combined Modality Therapy instrumentation, Combined Modality Therapy methods, Electrodes, Implanted, Female, Humans, Prosthesis Failure, Prosthesis Implantation instrumentation, Treatment Outcome, Atrial Fibrillation prevention & control, Atrioventricular Block prevention & control, Cardiac Catheterization methods, Pacemaker, Artificial, Prosthesis Implantation methods
- Abstract
An 83-year-old lady had a DDDR pacemaker inserted in 1997 for symptomatic atrioventricular block. She underwent battery replacement in 2008. In 2010, she developed atrial fibrillation; the pacemaker was switched to VVIR mode. During the last 2years, ventricular lead threshold increased progressively. In December 2015, she presented for elective battery replacement. After successful battery replacement, the ventricular lead threshold remained high; therefore, we implanted a leadless transcatheter pacemaker, via femoral vein access, using a dedicated catheter delivery system. Electrical measurements at this stage revealed a pacing threshold of 0.28V at 0.24msec, and an impedance of 650Ω., (Copyright © 2016 Elsevier Inc. All rights reserved.)