Objective: To investigate whether a solution of oxybuprocaine hydrochloride, 0.4%, results in a false-positive response in an immunochromatographic SAS Adeno Test., Design: Experimental study., Controls: Physiologic saline and 2% lidocaine., Testing: Each chemical (100 microl) was diluted in a transport medium. Five drops (200 microl) of the resultant solution were dispensed into the round sample well of a test device. Fifteen samples were tested in each group., Main Outcome Measures: Ten minutes after the start of the test, a colored line in the "specimen" portion of the test membrane was visually read as positive or negative by a masked technician., Results: No positive reaction was observed in the control groups (physiologic saline and lidocaine). A false-positive reaction was observed in six samples (33.3%) in the oxybuprocaine group. The positive rate was significantly higher in the oxybuprocaine group compared with those in control groups (P = 0.0062, Fisher's extract probability test)., Conclusions: Oxybuprocaine may induce a false-positive reaction in an immunochromatographic SAS Adeno Test. We recommend the use of lidocaine, instead of oxybuprocaine, for local anesthesia in taking eye swabs from patients with suspected adenovirus infection.