1. Underestimation of 68Ga PET/CT SUV caused by activity overestimation using default calibrator settings
- Author
-
Tom Sanderson, James Solomon, John Dickson, and Chris Nottage
- Subjects
PET-CT ,Scanner ,medicine.diagnostic_test ,business.industry ,Biophysics ,General Physics and Astronomy ,Computed tomography ,General Medicine ,Imaging phantom ,030218 nuclear medicine & medical imaging ,03 medical and health sciences ,0302 clinical medicine ,030220 oncology & carcinogenesis ,LOCAL TOLERANCE ,medicine ,Calibration ,Radiology, Nuclear Medicine and imaging ,Ct imaging ,Nuclear medicine ,business - Abstract
Purpose A PET/CT scan of a uniformly filled 68Ga phantom resulted in an unexpectedly low SUVmean of 0.88. A potential contributing cause of underestimation of 68Ga SUV is overestimation of 68Ga activity in the Radionuclide Calibrator associated with the PET/CT scanner. To investigate this, a Radionuclide Calibrator cross-calibration exercise was performed. Methods A source of 68Ga was measured in 5 Capintec CRC-55tR calibrators using the pre-set 68Ga calibrator factor of 416, and a Veenstra VDC-505 calibrator using a factor recommended by the manufacturer. The source was then measured in an externally located Fidelis Secondary Standard Radionuclide Calibrator. Manual adjustments were made to the Capintec calibrator factors to match the decay corrected Fidelis measurement, followed by a repeat PET/CT scan of a uniform 68Ga phantom. Results The cross-calibration results showed that the 5 Capintec calibrators systematically overestimated 68Ga activity by 7.8–9.4% (mean 8.5%) compared to the Fidelis. The calibrator factors were adjusted to 456–464 to match the Fidelis measurement, and the repeat phantom scan resulted in a SUVmean of 0.97, within the local tolerance of 1.00 ± 5%. The result for the Veenstra calibrator was within the tolerance of ±5%. Conclusions Underestimation of 68Ga SUV was primarily caused by overestimation of 68Ga activity using the pre-set calibrator factor setting on a Capintec CRC-55tR. Improvement in quantification accuracy was achieved by adjusting the 68Ga calibrator factor based upon a cross-calibration exercise. We recommend SUV checks using a uniform phantom and regular calibrator cross-calibration exercises for all isotopes used for quantitative PET/CT imaging.
- Published
- 2019
- Full Text
- View/download PDF