1. Validation of the iHealth Track upper-arm blood pressure monitor KN-550BT in general population according to the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018+AMD1:2020).
- Author
-
Shi L, Zhang XY, Sun JY, Chen JF, Ma YF, and Zhou KY
- Subjects
- Humans, Blood Pressure, Blood Pressure Determination, Systole, Adenosylmethionine Decarboxylase, Blood Pressure Monitors, Hypertension
- Abstract
Objective: To validate the iHealth Track KN-550BT oscillometric upper-arm blood pressure monitor in general population according to the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018+AMD1:2020)., Methods: Participants were recruited and the same left-arm sequential method was used for blood pressure measurement according to the ISO 81060-2:2018+AMD1:2020. The validation results were assessed following the protocol and the Bland-Altman scatterplot was used to show the difference between the test device and reference results., Results: A total of 89 qualified participants were included in the final analysis. For the validation Criterion 1, the mean ± SD of the differences between the test device and reference readings was -1.22 ± 5.76 mmHg and -0.08 ± 4.40 mmHg for systolic and diastolic blood pressure, respectively. For Criterion 2, the mean ± SD of the differences between the test device and reference readings per participant was -1.22 ± 5.06 mmHg and -0.08 ± 3.84 mmHg for systolic and diastolic blood pressure, respectively., Conclusion: The iHealth Track KN-550BT upper-arm blood pressure monitor passed all the requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018+AMD1:2020) and can be recommended for clinical use and self-measurement in general population., (Copyright © 2023 Wolters Kluwer Health, Inc. All rights reserved.)
- Published
- 2024
- Full Text
- View/download PDF