1. Comparison of polypropylene and silicone Ahmed® glaucoma valves in the treatment of neovascular glaucoma: A 2-year follow-up.
- Author
-
Lubiński W, Krzystolik K, Gosławski W, Kuprjanowicz L, and Mularczyk M
- Subjects
- Follow-Up Studies, Humans, Intraocular Pressure physiology, Postoperative Complications, Retrospective Studies, Treatment Outcome, Glaucoma Drainage Implants, Glaucoma, Neovascular surgery, Polypropylenes, Prosthesis Implantation methods, Silicones
- Abstract
Background: Inflammation associated with biomaterials of Ahmed® glaucoma drainage devices may cause the formation of a capsule around the device and can thus have a significant influence on the level of intraocular pressure reduction., Objectives: The objective of this study was to compare the clinical outcomes after the implantation of a polypropylene or silicone Ahmed® glaucoma valve in patients with neovascular glaucoma., Material and Methods: In the study, 27 eyes with neovascular glaucoma (group 1) received silicon Ahmed® valves and 23 eyes (group 2) received polypropylene valves. The best corrected distance visual acuity (BCDVA), intraocular pressure (IOP) and number of anti-glaucomatous drugs were recorded preoperatively and during a follow-up period of 24 months after surgery. Success was defined by the following criteria: 1) intraocular pressure in the rage of 6-21 mm Hg; 2) IOP reduction of at least 30% relative to preoperative values. All complications were registered., Results: One month postoperatively, the mean BCDVA increased significantly in both groups compared to preoperative values (p < 0.001). These values did not change during the 24 months of follow-up examinations. The probability of success defined by criterion 1 at 24 months of observation was 66.7% for silicone and 27.3% for propylene valves group (p < 0.007). According to criterion 2, the difference in success between the groups was not statistically significant. The total number of complications that occurred in both groups during the 24 months of follow-up examinations was similar, except for a higher occurrence of Tenon's cyst formation in the group with a polypropylene valve (18% vs 35%; p < 0.04)., Conclusions: In patients with neovascular glaucoma, the implantation of a silicone valve is associated with a significantly higher probability of long-term reduction of IOP below 21 mm Hg and with a lower risk of valve encapsulation in comparison to polypropylene valves. The obtained results suggest that silicone Ahmed® valves are more effective in the treatment of patients with neovascular glaucoma.
- Published
- 2018
- Full Text
- View/download PDF