1. A Novel Implant Design for Cemented Endoprosthesis Stems to Reduce the Risk of Bone Cement Implantation Syndrome.
- Author
-
Udogwu UN, Fogel JD, Sim D, Sahlani MN, Smith R, Aroom KR, and Ng VY
- Abstract
Background: Bone cement implantation syndrome (BCIS) is a life-threatening risk of cemented stems. There are limited methods to prevent BCIS and few published studies. A survey of the Musculoskeletal Tumor Society (MSTS) surgeons was conducted to evaluate their experiences with BCIS. A novel stem designed to reduce intramedullary pressure during insertion was evaluated., Methods: A survey was distributed to MSTS members, and data was collected. The novel stem featured a hollow longitudinal channel, an entry hole at the tip, and an egress hole at the collar for cement to flow from the intramedullary canal during insertion. Bending stiffness was compared using finite element analysis to a standard solid stem. Stems were cemented into cadaveric femurs. Specimens were loaded with 8000 N tensile force and then maximally torqued until failure. Intramedullary pressures were measured for novel and standard stems cemented into sawbones., Results: In 107 survey responses, 58% (n = 63) experienced severe BCIS, and 83% (n = 52) of those had ≥1 death from BCIS complications. Many surgeons avoid cementing long stems, and 78% (n = 14) report concern for BCIS as the reason. Seventy-nine percent (n = 84) use an average of 4.75 different methods to reduce BCIS risk. The novel stem demonstrated 2.8% reduced bending stiffness. When cemented into cadaveric bone, both stem designs achieved 8000 N of tensile force, and there was no significant difference in torque failure (140.6 Nm in the novel stem; 128 Nm in the standard stem). The average peak pressure was significantly lower for the novel stem (77 psi vs. 151 psi)., Conclusions: The majority of surgeons have concerns regarding BCIS based on their experience and use multiple methods to reduce risk. A novel stem can reduce the intramedullary insertion pressure by approximately half and possibly reduce the risk of BCIS. Testing demonstrates similar stiffness and stability compared to standard solid stems., Competing Interests: Human subjects: Consent for treatment and open access publication was obtained or waived by all participants in this study. N/a issued approval n/a. IRB approval was waived as the survey portion of this experiment was completely anonymous and no identifiable information was obtained from the participants. Consent for the survey was implied by respondent participation. Animal subjects: All authors have confirmed that this study did not involve animal subjects or tissue. Conflicts of interest: In compliance with the ICMJE uniform disclosure form, all authors declare the following: Payment/services info: All authors have declared that no financial support was received from any organization for the submitted work. Financial relationships: All authors have declared that they have no financial relationships at present or within the previous three years with any organizations that might have an interest in the submitted work. Intellectual property info: The novel hip stem Safe Stem is a patented product. Patent number: US12070392B2. Other relationships: All authors have declared that there are no other relationships or activities that could appear to have influenced the submitted work., (Copyright © 2025, Udogwu et al.)
- Published
- 2025
- Full Text
- View/download PDF