1. The potential of microcalorimeter x-ray spectrometers for measurement of relative fluorescence-line intensities
- Author
-
Fowler, J. W., Miaja-Avila, L., O'Neil, G. C., Ullom, J. N., Whitelock, H., and Swetz, D. S.
- Subjects
Physics - Instrumentation and Detectors ,Physics - Atomic Physics - Abstract
We have previously used an array of cryogenic microcalorimeters with 4 eV energy resolution to measure emission-line profiles and energies of the characteristic L-shell x rays of four elements of the lanthanide series: praseodymium, neodymium, terbium, and holmium. We consider the power of the same data set for the estimation of the lines' relative intensities. Intensities must be corrected for detector efficiency and self-absorption, and we estimate uncertainties on the corrections. These data represent one of the first uses of cryogenic energy-dispersive sensors to estimate the relative intensities of x-ray fluorescence lines. They show that a future measurement of thin-film samples with microcalorimeter detectors could achieve systematic uncertainties below 1% on relative line intensities over a broad energy range., Comment: Submitted to Radiation Physics & Chemistry, special issue on x-ray fundamental parameters
- Published
- 2022
- Full Text