1. Predictors of increase in pacing threshold after transcatheter pacing system implantation due to micro-dislodgement
- Author
-
Yoshiro Chiba, Kazutaka Aonuma, Minoru Murata, Yasutoshi Shinoda, Koji Ohira, Norihiro Yamada, Tomoaki Hasegawa, Mayu Ishibashi, and Motoaki Higuchi
- Subjects
Male ,medicine.medical_specialty ,Cardiac Catheterization ,Pacemaker, Artificial ,Miniaturization ,business.industry ,Cardiac Pacing, Artificial ,Arrhythmias, Cardiac ,General Medicine ,Equipment Design ,030204 cardiovascular system & hematology ,Equipment Failure Analysis ,03 medical and health sciences ,0302 clinical medicine ,Electric Power Supplies ,Internal medicine ,medicine ,Cardiology ,Humans ,Female ,030212 general & internal medicine ,Implant ,Cardiology and Cardiovascular Medicine ,business ,Aged - Abstract
Achieving a favorable pacing threshold with a Micra transcatheter pacing system (Micra-TPS) is needed to reduce battery depletion. In some cases, the threshold increases shortly after the device is implanted, and a higher pacing threshold may be required. This study aims to identify the causes and predictors of the increase in pacing threshold observed shortly after Micra-TPS implantation.The study included 64 consecutive patients who underwent Micra-TPS implantation between 2017 and 2020. The patients were divided into two groups depending on their pacing threshold: the increased pacing threshold (IPT) group (threshold increased by ≥0.5 V/0.24 ms within 1 month of implantation) and the stable pacing threshold (SPT) group.Excluding four patients who could not be followed up, of the 60 remaining patients, nine (15%) were in the IPT group and 51 (85%) were in the SPT group. The IPT group had significantly lower implant impedance values and higher implant thresholds than the SPT group (582 ± 59 vs 755 ± 167 Ω [P .001] and 1.29 ± 0.87 vs 0.71 ± 0.40 V/0.24 ms [P = .014]). Implant impedance and threshold may serve as predictors of a threshold increase after implantation (area under the curve: 0.737-0.943 and 0.586-0.926, respectively).An IPT was noted shortly after Micra-TPS implantation owing to micro-dislodgement because of insufficient anchoring of the device to the myocardium. Impedance 660 Ω and threshold 1.0 V/0.24 ms may predict an increase in pacing threshold.
- Published
- 2020