1. Twelve-Week 24/7 Ambulatory Artificial Pancreas With Weekly Adaptation of Insulin Delivery Settings: Effect on Hemoglobin A 1c and Hypoglycemia.
- Author
-
Dassau E, Pinsker JE, Kudva YC, Brown SA, Gondhalekar R, Dalla Man C, Patek S, Schiavon M, Dadlani V, Dasanayake I, Church MM, Carter RE, Bevier WC, Huyett LM, Hughes J, Anderson S, Lv D, Schertz E, Emory E, McCrady-Spitzer SK, Jean T, Bradley PK, Hinshaw L, Laguna Sanz AJ, Basu A, Kovatchev B, Cobelli C, and Doyle FJ 3rd
- Subjects
- Adult, Blood Glucose, Blood Glucose Self-Monitoring, Female, Humans, Hypoglycemia drug therapy, Insulin Infusion Systems, Male, Middle Aged, Pancreas, Artificial, Diabetes Mellitus, Type 1 drug therapy, Glycated Hemoglobin metabolism, Hypoglycemic Agents administration & dosage, Insulin administration & dosage
- Abstract
Objective: Artificial pancreas (AP) systems are best positioned for optimal treatment of type 1 diabetes (T1D) and are currently being tested in outpatient clinical trials. Our consortium developed and tested a novel adaptive AP in an outpatient, single-arm, uncontrolled multicenter clinical trial lasting 12 weeks., Research Design and Methods: Thirty adults with T1D completed a continuous glucose monitor (CGM)-augmented 1-week sensor-augmented pump (SAP) period. After the AP was started, basal insulin delivery settings used by the AP for initialization were adapted weekly, and carbohydrate ratios were adapted every 4 weeks by an algorithm running on a cloud-based server, with automatic data upload from devices. Adaptations were reviewed by expert study clinicians and patients. The primary end point was change in hemoglobin A
1c (HbA1c ). Outcomes are reported adhering to consensus recommendations on reporting of AP trials., Results: Twenty-nine patients completed the trial. HbA1c , 7.0 ± 0.8% at the start of AP use, improved to 6.7 ± 0.6% after 12 weeks (-0.3, 95% CI -0.5 to -0.2, P < 0.001). Compared with the SAP run-in, CGM time spent in the hypoglycemic range improved during the day from 5.0 to 1.9% (-3.1, 95% CI -4.1 to -2.1, P < 0.001) and overnight from 4.1 to 1.1% (-3.1, 95% CI -4.2 to -1.9, P < 0.001). Whereas carbohydrate ratios were adapted to a larger extent initially with minimal changes thereafter, basal insulin was adapted throughout. Approximately 10% of adaptation recommendations were manually overridden. There were no protocol-related serious adverse events., Conclusions: Use of our novel adaptive AP yielded significant reductions in HbA1c and hypoglycemia., (© 2017 by the American Diabetes Association.)- Published
- 2017
- Full Text
- View/download PDF