1. Feasibility of using micro silica bead TLDs for in-Vivo dosimetry of CT-based HDR prostate brachytherapy: An experimental and simulation study.
- Author
-
Siavashpour Z, Kiarad R, Aghamiri MR, Babaloui S, Seiri M, and Jaberi R
- Subjects
- Male, Humans, Tomography, X-Ray Computed, Radiotherapy Dosage, Radiotherapy Planning, Computer-Assisted methods, Thermoluminescent Dosimetry methods, Thermoluminescent Dosimetry instrumentation, In Vivo Dosimetry methods, Brachytherapy methods, Brachytherapy instrumentation, Silicon Dioxide chemistry, Prostatic Neoplasms radiotherapy, Monte Carlo Method, Feasibility Studies, Phantoms, Imaging
- Abstract
Purpose: Feasibility of silica-based dosimeters for IVD of HDR prostate brachytherapy., Material and Methods: Plastic dosimeter holders and a water-fillable prostate phantom were built in-house. Interstitial prostate brachytherapy and Monte Carlo simulations were performed. The treatment planning, Monte-Carlo simulation, and dosimetry results were compared., Results: The relative differences between TLD-TPS, TLD-MCNP, and TPS-MCNP were 0.2-6.9 %, 0.5-6.5 %, and 0.6-6.3 %, respectively., Conclusion: Micro-silica bead dosimeters can perform offline in situ quality assurance in HDR prostate brachytherapy., Competing Interests: Declaration of competing interest The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper., (Copyright © 2024 Elsevier Ltd. All rights reserved.)
- Published
- 2024
- Full Text
- View/download PDF