1. [The influence of viscosity of synthetic urine on the flow rate through catheters].
- Author
-
Braunwarth H, Brill FH, Hegeholz D, and Hammerer P
- Subjects
- Biomechanical Phenomena, Cellulose analogs & derivatives, Equipment Design, Humans, In Vitro Techniques, Time Factors, Viscosity, Urinary Catheterization instrumentation, Urine chemistry, Urine physiology, Urodynamics physiology
- Abstract
Background: There are special requirements for draining a pouch, e.g. length, flexibility and ability to drain highly viscous urine., Methods: We checked in vitro the flow time of pure and synthetic urine thickened with hydroxyethylcellulose (0.125% and 0.250%, HEC) on selected catheters., Results: We found for SpeediCath Compact Male (SCCM) as a CH12 catheter the shortest flow times for undiluted and urine thickened with 0.125% and for SafetyCath Plus for urine thickened with 0.250%. For catheters with CH14 we found significantly shorter flow times for all tested solutions. The flow time does not depend on amount and area of the catheter eyes; we found no difference in the flow time for 4-eye (MobiStom) and 2-eye catheters.
- Published
- 2012