1. Valve-in-Valve Implantation Using the ACURATE Neo in Degenerated Aortic Bioprostheses: An International Multicenter Analysis.
- Author
-
Holzamer A, Kim WK, Rück A, Sathananthan J, Keller L, Cosma J, Bauer T, Nef H, Amat-Santos IJ, Brinkert M, Husser O, Pellegrini C, Schofer J, Nerla R, Montorfano M, Giannini F, Stella P, Kuwata S, Hilker M, Castriota F, Ussia GP, Webb JG, Nietlispach F, and Toggweiler S
- Subjects
- Aged, Aged, 80 and over, Aortic Valve diagnostic imaging, Aortic Valve physiopathology, Canada, Europe, Female, Heart Valve Prosthesis Implantation adverse effects, Hemodynamics, Humans, Male, Recovery of Function, Retrospective Studies, Risk Factors, Time Factors, Transcatheter Aortic Valve Replacement adverse effects, Treatment Outcome, Aortic Valve surgery, Bioprosthesis, Heart Valve Prosthesis, Heart Valve Prosthesis Implantation instrumentation, Prosthesis Failure, Stents, Transcatheter Aortic Valve Replacement instrumentation
- Abstract
Objectives: This study reports an international experience using the transfemoral ACURATE neo transcatheter heart valve (Boston Scientific, Marlborough, Massachusetts) for the treatment of degenerated surgical aortic bioprostheses., Background: Transcatheter valve-in-valve procedures have emerged as an alternative to redo surgery. Supra-annular prostheses might be particularly useful in this indication., Methods: This is an international multicenter analysis including 85 patients from 14 centers in Europe and Canada undergoing an ACURATE neo valve-in-valve procedure from March 2015 to February 2019., Results: Internal diameter of the degenerated bioprosthesis was 20.3 ± 2.1 mm. Prosthesis size S was used in 70 (82%) procedures. The median depth of implantation was 3 mm and the upper crown of the ACURATE neo was positioned above the stent posts of the degenerated bioprosthesis in 54 (64%) and inside in 31 (36%). Mean transvalvular gradient before discharge was significantly lower if the upper crown was above the degenerated bioprosthesis (13.7 ± 5.9 mm Hg vs. 19.5 ± 10.0 mm Hg; p = 0.001). However, a high position of the ACURATE neo resulted in embolization in 1 patient, conversion to open-heart surgery in 1, and need for reintervention due to transcatheter heart valve failure within the first 18 months of follow-up in 4., Conclusions: This early experience shows that a high implantation of the ACURATE neo with the upper crown above the stent posts of the degenerated bioprosthesis resulted in lower mean transvalvular gradients but a higher rate of malpositioning and early valve degeneration., (Copyright © 2019 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.)
- Published
- 2019
- Full Text
- View/download PDF