1. Design and implementation of three-dimensional ring-scanning equipment for optimized measurements of near-infrared diffuse optical breast imaging.
- Author
-
Jhao-Ming Yu, Min-Cheng Pan, Ya-Fen Hsu, Liang-Yu Chen, and Min-Chun Pan
- Subjects
- *
BREAST imaging , *DIAGNOSTIC imaging , *NEAR infrared radiation , *THREE-dimensional imaging , *OPTICAL resolution - Abstract
We propose and implement three-dimensional (3-D) ring-scanning equipment for near-infrared (NIR) diffuse optical imaging to screen breast tumors under prostrating examination. This equipment has the function of the radial, circular, and vertical motion without compression of breast tissue, thereby achieving 3-D scanning; furthermore, a flexible combination of illumination and detection can be configured for the required resolution. Especially, a rotation-sliding-and-moving mechanism was designed for the guidance of source- and detectionchannel motion. Prior to machining and construction of the system, a synthesized image reconstruction was simulated to show the feasibility of this 3-D NIR ring-scanning equipment; finally, this equipment is verified by performing phantom experiments. Rather than the fixed configuration, this addressed screening/diagnosing equipment has the flexibilities of optical-channel expansion for spatial resolution and the dimensional freedom for scanning in reconstructing optical-property images. [ABSTRACT FROM AUTHOR]
- Published
- 2015
- Full Text
- View/download PDF