1. Multiphase preclinical assessment of a novel device to locate unintentionally retained surgical sharps: a proof-of-concept study
- Author
-
Hae Sung Kang, Jad Khoraki, Jessie Gie, Dielle Duval, Susan Haynes, Michael Siev, Jay Shah, Fernando Kim, Martin Mangino, Levi Procter, Riccardo Autorino, and Samuel Weprin
- Subjects
Patient safety ,Retained surgical sharps ,Melzi Sharps Finder ,Never event ,Surgery ,RD1-811 - Abstract
Abstract Background Retained surgical sharps (RSS) is a “never event” that is preventable but may still occur despite of correct count and negative X-ray. This study assesses the feasibility of a novel device (“Melzi Sharps Finder®” or MSF) in effective detection of RSS. Methods The first study consisted of determination of the presence of RSS or identification of RSS in an ex-vivo model (a container with hay in a laparoscopic trainer box). The second study consisted of determining presence of RSS in an in-vivo model (laparoscopy in live adult Yorkshire pigs) with 3 groups: C-arm, C-arm with MSF and MSF. The third study used similar apparatus though with laparotomy and included 2 groups: manual search and MSF. Results In the first study, the MSF group had a higher rate of identification of a needle and decreased time to locate a needle versus control (98.1% vs. 22.0%, p
- Published
- 2023
- Full Text
- View/download PDF