1. Commissioning of the double Penning trap system MLLTRAP
- Author
-
J. B. Neumayr, Dietrich Habs, Michael Bussmann, J. Szerypo, E. Gartzke, V. S. Kolhinen, C. Schürmann, and P. G. Thirolf
- Subjects
Physics ,Nuclear and High Energy Physics ,Mass separation ,Penning trap ,Mass measurement ,Ion ,Trap (computing) ,Nuclear physics ,Buffer gas cooling ,Relative mass ,Ion trap ,Atomic physics ,Instrumentation - Abstract
A cylindrical double Penning trap system has been installed and commissioned at the Maier-Leibnitz-Laboratory (MLL) in Garching. This trap system has been designed to isobarically purify low-energy ion beams and perform highly accurate mass measurements. Technical details of the device and the first results of the commissioning measurements will be presented. The mass resolving power achieved in the first trap for 85 Rb ions is R = 139 ( 2 ) × 10 3 , while a relative mass uncertainty of δ m / m = 2.9 × 10 − 8 was reached with the second trap (no analysis of systematic uncertainties included) when using 87 Rb as a reference ion for 85 Rb.
- Published
- 2009
- Full Text
- View/download PDF