Back to Search
Start Over
Determination of ultratrace-levels of Tc in low-level radioactive waste samples using ICP-QMS.
- Source :
-
Journal of Radioanalytical & Nuclear Chemistry . Feb2015, Vol. 303 Issue 2, p1245-1248. 4p. - Publication Year :
- 2015
-
Abstract
- This paper proposes a rapid, accurate method of measuring Tc in low-level radioactive waste using microwave digestion, TEVA resin for chemical separation, and inductively coupled plasma-quadrupole mass spectrometry (ICP-QMS). The minimum detectable activity of ICP-QMS for the determination of Tc was 8.5 mBq g (13.6 pg g). The proposed method achieved chemical recovery of 88 % and with sensitivity superior to that of alternative radiochemical methods (e.g., liquid scintillation analyzer or LSA) when the data acquisition time for identical, low-concentration samples was taken into account. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 02365731
- Volume :
- 303
- Issue :
- 2
- Database :
- Academic Search Index
- Journal :
- Journal of Radioanalytical & Nuclear Chemistry
- Publication Type :
- Academic Journal
- Accession number :
- 100489990
- Full Text :
- https://doi.org/10.1007/s10967-014-3490-6