Validation of Method Performance of pH, PCO2, PO2, Na+, K+ of Cobas b121 ABG Analyser.

Authors :
KUMAR NANDA, SUNIL
RAY, LOPAMUDRA
DINAKARAN, ASHA
Source :
Journal of Clinical & Diagnostic Research. Jun2014, Vol. 8 Issue 6, p5-7. 3p.
Publication Year :
2014

Abstract

Introduction: The introduction of a new method or new analyser is a common occurrence in clinical biochemistry laboratory. Blood gas measurements and electrolytes are often performed in Point-of-Care (POC) settings. When a new POC analyser is obtained, the performance of the analyser should be evaluated by comparison to the measurements with the reference analyser in the laboratory. Objectives: Evaluation of method performance of pH, PCO2, PO2, Na+, K+ of cobas b121 ABG analyser. Materials & Methods: The evaluation of method performance of pH, PO2, PCO2, Na+, K+ of cobas b121 ABG analyser was done by comparing the results of 50 patient samples run on cobas b121 with the results obtained from Rapid lab values (reference analyser). Correlation coefficient was calculated from the results obtained from both the analysers. Precision was calculated by running biorad ABG control samples. Results: The correlation coefficient values obtained for parameters were close to 1.0 indicating good correlation. The CV obtained for all the parameters were less than 5 indicating good precision. Conclusion: The new ABG analyser, Cobas b121 correlated well with the reference ABG analyser (Rapid Lab) and could be used to run on patient samples. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
0973709X
Volume :
8
Issue :
6
Database :
Academic Search Index
Journal :
Journal of Clinical & Diagnostic Research
Publication Type :
Academic Journal
Accession number :
101625283
Full Text :
https://doi.org/10.7860/JCDR/2014/9218.4442