Back to Search Start Over

Sampling Current Ratio Measurement System for Calibration of Current Transducers up to 10 kA With 5\cdot 10^\mathrm -6 Uncertainty.

Authors :
van den Brom, Helko E.
Rietveld, Gert
So, Eddy
Source :
IEEE Transactions on Instrumentation & Measurement. Jun2015, Vol. 64 Issue 6, p1685-1691. 7p.
Publication Year :
2015

Abstract

A sampling current ratio measurement system has been developed, tested, and validated for the accurate complex ratio measurement of ac current transducers. Calibrations can be performed for primary currents up to 10 kA with typical expanded uncertainties of 5\cdot 10^\mathrm \mathbf -6 in magnitude and 5 $\mu $ rad in phase ( $k=2$ ). A unique property of the system is the use of digitizers for sampling of the secondary current signals and step-down transformers with a large number of current ratios. Therefore, the system does not require the current transducer under test to be a current transformer having a nominal ratio equal to the reference current transformer. Furthermore, it allows for the accurate calibration of a large variety of different current transformers and other types of current transducers. A bilateral comparison between VSL and NRC confirms the stated uncertainty. [ABSTRACT FROM PUBLISHER]

Details

Language :
English
ISSN :
00189456
Volume :
64
Issue :
6
Database :
Academic Search Index
Journal :
IEEE Transactions on Instrumentation & Measurement
Publication Type :
Academic Journal
Accession number :
102615355
Full Text :