An attempt to create a standardized (reference) model for experimental investigations on implant’s sample.

Authors :
Tanasić, Ivan
Tihaček-Šojić, Ljiljana
Mitrović, Nenad
Milić-Lemić, Aleksandra
Vukadinović, Miroslav
Marković, Aleksa
Milošević, Miloš
Source :
Measurement (02632241). Aug2015, Vol. 72, p37-42. 6p.
Publication Year :
2015

Abstract

The aim of this study was to determine, evaluate and measure strain of vertically loaded implant using the digital image correlation method. The Straumann® dental implant system with the SLActive® surface was used in this study. Implant was immersed in poly-methyl-methacrylate during his hardening process. After preparation procedure a sample of implant and poly-methyl-methacrylate was obtained. This sample was loaded using external load from 0 to 400 N. Maximum strain in the 4 mm surface layer was 0.30% whereas maximum strain in the 6 mm surface layer (opposite side) was 0.20%, detected in the marginal and apical part of implant-poly-methyl-methacrylate sample. Minimum strain measured by Aramis software was 0.01%, detected in the 4 mm surface layer. According to results obtained by Aramis data processing, the 4 mm surface layer indicated higher overall strain in apical direction with the strains of 0.18–0.21%. Increasing the load did not affect the value of maximum strain however, higher load influenced the overall strain concentration increased especially in marginal and apical part of the sample surfaces that surrounded lateral sides of implant body. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
02632241
Volume :
72
Database :
Academic Search Index
Journal :
Measurement (02632241)
Publication Type :
Academic Journal
Accession number :
103117190
Full Text :
https://doi.org/10.1016/j.measurement.2015.04.029