Back to Search Start Over

Influence of skin blood flow and source-detector distance on near-infrared spectroscopy-determined cerebral oxygenation in humans.

Authors :
Hirasawa, Ai
Yanagisawa, Shintaro
Tanaka, Naoki
Funane, Tsukasa
Kiguchi, Masashi
Sørensen, Henrik
Secher, Niels H.
Ogoh, Shigehiko
Source :
Clinical Physiology & Functional Imaging. May2015, Vol. 35 Issue 3, p237-244. 8p. 1 Diagram, 1 Chart, 4 Graphs.
Publication Year :
2015

Abstract

Most near-infrared spectroscopy ( NIRS) apparatus fails to isolate cerebral oxygenation from an extracranial contribution although they use different source-detector distances. Nevertheless, the effect of different source-detector distances and change in extracranial blood flow on the NIRS signal has not been identified in humans. This study evaluated the extracranial contribution, as indicated by forehead skin blood flow (Sk BF) to changes in the NIRS-determined cerebral oxyhaemoglobin concentration (O2Hb) by use of a custom-made multidistance probe. Seven males (age 21 ± 1 year) were in a semi-recumbent position, while extracranial blood flow was restricted by application of four different pressures (+20 to +80 mmHg) to the left temporal artery. The O2Hb was measured at the forehead via a multidistance probe (source-detector distance; 15, 22·5 and 30 mm), and Sk BF was determined by laser Doppler. Heart rate and blood pressure were unaffected by application of pressure to the temporal artery, while Sk BF gradually decreased ( P<0·001), indicating that extracranial blood flow was manipulated without haemodynamic changes. Also, O2Hb gradually decreased with increasing applied pressure ( P<0·05), and the decrease was related to that in Sk BF ( r = 0·737, P<0·01) independent of the NIRS source to detector distance. These findings suggest that the NIRS-determined cerebral oxyhaemoglobin is affected by change in extracranial blood flow independent of the source-detector distance from 15 to 30 mm. Therefore, new algorithms need to be developed for unbiased NIRS detection of cerebral oxygenation. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
14750961
Volume :
35
Issue :
3
Database :
Academic Search Index
Journal :
Clinical Physiology & Functional Imaging
Publication Type :
Academic Journal
Accession number :
103394891
Full Text :
https://doi.org/10.1111/cpf.12156