Robot arm based flat panel CT-guided electromagnetic tracked spine interventions: phantom and animal model experiments.
- Source :
-
European Radiology . Nov2010, Vol. 20 Issue 11, p2656-2662. 7p. - Publication Year :
- 2010
-
Abstract
- <bold>Purpose: </bold>To evaluate accuracy and procedure times of electromagnetic tracking (EMT) in a robotic arm mounted flat panel setting using phantom and animal cadaveric models.<bold>Methods and Materials: </bold>A robotic arm mounted flat panel (RMFP) was used in combination with EMT to perform anthropomorphic phantom (n = 90) and ex vivo pig based punctures (n = 120) of lumbar facet joints (FJ, n = 120) and intervertebral discs (IVD, n = 90). Procedure accuracies and times were assessed and evaluated.<bold>Results: </bold>FJ punctures were carried out with a spatial accuracy of 0.8 ± 0.9 mm (phantom) and 0.6 ± 0.8 mm (ex vivo) respectively. While IVD punctures showed puncture deviations of 0.6 ± 1.2 mm (phantom) and 0.5 ± 0.6 mm (ex vivo), direct and angulated phantom based punctures had accuracies of 0.8 ± 0.9 mm and 1.0 ± 1.3 mm. Planning took longer for ex vivo IVD punctures compared to phantom model interventions (39.3 ± 17.3 s vs. 20.8 ± 5.0 s, p = 0.001) and for angulated vs. direct phantom FJ punctures (19.7 ± 5.1 s vs. 28.6 ± 7.8 s, p < 0.001). Puncture times were longer for ex vivo procedures when compared to phantom model procedures in both FJ (37.9 ± 9.0 s vs. 23.6 ± 7.2 s, p = 0.001) and IVD punctures (43.9 ± 16.1 s vs. 31.1 ± 6.4 s, p = 0.026).<bold>Conclusion: </bold>The combination of RMFP with EMT provides an accurate method of navigation for spinal interventions such as facet joint punctures and intervertebral disc punctures. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 09387994
- Volume :
- 20
- Issue :
- 11
- Database :
- Academic Search Index
- Journal :
- European Radiology
- Publication Type :
- Academic Journal
- Accession number :
- 104927338
- Full Text :
- https://doi.org/10.1007/s00330-010-1837-0