Back to Search Start Over

Hydrodynamic characteristics of a membrane oxygenator: modeling of pressure-flow characteristics and their influence on apparent viscosity.

Authors :
Okahara, S.
Tsuji, T.
Ninomiya, S.
Miyamoto, S.
Takahashi, H.
Soh, Z.
Sueda, T.
Source :
Perfusion. Sep2015, Vol. 30 Issue 6, p478-483. 6p. 1 Diagram, 3 Charts, 2 Graphs.
Publication Year :
2015

Abstract

The viscosity obtained from pressure-flow characteristics of an oxygenator may help to detect factors that change oxygenator resistance. The objective of this study was to model pressure-flow characteristics of a membrane oxygenator with an integrated arterial filter and to quantify their influence on apparent viscosity of non-Newtonian fluids. One Newtonian fluid (glycerin solution) and two non-Newtonian fluids (whole bovine blood and a human red blood cell suspension) were perfused through an oxygenator and their pressure-flow characteristics examined systematically. Four resistance parameters for the pressure gradient characteristics approximation equation were obtained by the least squares method from the relational expression of pressure-flow characteristics and viscosity. For all three fluids, a non-linear flow to pressure change was observed with a coefficient of determination of almost 1 by exponential approximation. The glycerin solution had a higher pressure gradient (10-70%) than the other fluids; the apparent viscosity of the non-Newtonian fluids was around 35% lower than the static one measured by a torsional oscillation viscometer. Overall, our study demonstrated that the influence on the apparent viscosity of non-Newtonian fluids can be quantified by pressure gradient differences in a membrane oxygenator with an integrated arterial filter. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
02676591
Volume :
30
Issue :
6
Database :
Academic Search Index
Journal :
Perfusion
Publication Type :
Academic Journal
Accession number :
108996877
Full Text :
https://doi.org/10.1177/0267659114562101