Development of A phantom for ophthalmic beta source applicator quality control using TL dosimetry.

Authors :
Barbosa, N.A.
da Rosa, L.A.R.
Braz, D.
Source :
Radiation Physics & Chemistry. Nov2015, Vol. 116, p287-291. 5p.
Publication Year :
2015

Abstract

Concave eye applicators with 90 Sr/ 90 Y and 106 Ru/ 106 Rh beta ray sources are usually used in brachytherapy for the treatment of superficial intraocular tumors as uveal melanoma with thickness up to 5 mm. The calculation of the dose delivered to the eye is carried out based on the data present in the beta source calibration certificate. Therefore, it would be interesting to have a system that could evaluate that dose. In this work, an eye phantom to be used with 106 Ru/ 106 Rh betatherapy applicators was developed in solid water. This phantom can hold nine micro-cube thermoluminescent (TL) dosimeters, TLD-100. The characteristics of the TL response of the dosimeters, namely reproducibility and individual sensitivity, were determined for a 60 Co source. Using Monte Carlo code MCNPX, the dose to a water eye was determined at different depths. Exposing the eye phantom with TL dosimeters to the 106 Ru/ 106 Rh applicator, it is possible to assess calibration factors using the dose values obtained by Monte Carlo simulation to each depth. Using mean calibration factors, dose values obtained by TL dosimetry were compared to the data present in the applicators certificate. Mean differences for both applicators were lower than ±10%, maximum value 17% and minimum value 0.08%. Considering that the certificate values present an uncertainty of ±20%, the calibration procedure and the developed phantom are validated and can be applied. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
0969806X
Volume :
116
Database :
Academic Search Index
Journal :
Radiation Physics & Chemistry
Publication Type :
Academic Journal
Accession number :
109568908
Full Text :
https://doi.org/10.1016/j.radphyschem.2015.01.037