Back to Search Start Over

Noncontact diffuse correlation tomography of human breast tumor.

Authors :
Lian He
Yu Lin
Chong Huang
Irwin, Daniel
Szabunio, Margaret M.
Guoqiang Yu
Source :
Journal of Biomedical Optics. Aug2015, Vol. 20 Issue 8, p1-11. 11p.
Publication Year :
2015

Abstract

Our first step to adapt our recently developed noncontact diffuse correlation tomography (ncDCT) system for three-dimensional (3-D) imaging of blood flow distribution in human breast tumors is reported. A commercial 3-D camera was used to obtain breast surface geometry, which was then converted to a solid volume mesh. An ncDCT probe scanned over a region of interest on the mesh surface and the measured boundary data were combined with a finite element framework for 3-D image reconstruction of blood flow distribution. This technique was tested in computer simulations and in vivo human breasts with low-grade carcinoma. Results from computer simulations suggest that relatively high accuracy can be achieved when the entire tumor is within the sensitive region of diffuse light. Image reconstruction with a priori knowledge of the tumor volume and location can significantly improve the accuracy in recovery of tumor blood flow contrasts. In vivo imaging results from two breast carcinomas show higher average blood flow contrasts (5.9- and 10.9-fold) in the tumor regions compared to the surrounding tissues, which are comparable with previous findings using diffuse correlation spectroscopy. The ncDCT system has the potential to image blood flow distributions in soft and vulnerable tissues without distorting tissue hemodynamics. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
10833668
Volume :
20
Issue :
8
Database :
Academic Search Index
Journal :
Journal of Biomedical Optics
Publication Type :
Academic Journal
Accession number :
111803148
Full Text :
https://doi.org/10.1117/1.JBO.20.8.086003