Amperometric Determination of Sulfide by Glassy Carbon Electrode Modified with Hemin Functionalized Reduced Graphene Oxide.
- Source :
-
Electroanalysis . Jan2016, Vol. 28 Issue 1, p140-144. 5p. - Publication Year :
- 2016
-
Abstract
- This work describes a new approach for rapid detection of sulfide using a glassy carbon electrode (GCE) modified with hemin functionalized reduced graphene oxide (hemin-rGO). The modified GCE was applied to the selective amperometric detection of sulfide at a working potential of 0.05 V (vs. Ag/AgCl) at pH 7.5, which gave a linear range of 0.002-0.212 mM, a sensitivity of 246.5 µA mM−1 cm−2, and a detection limit of 1.3 µM. The proposed method was applied in determination of sulfide-spiked water, and the method was verified with recovery percentage of standard samples with an acceptable satisfaction. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 10400397
- Volume :
- 28
- Issue :
- 1
- Database :
- Academic Search Index
- Journal :
- Electroanalysis
- Publication Type :
- Academic Journal
- Accession number :
- 112234877
- Full Text :
- https://doi.org/10.1002/elan.201500508