Abstract
- A fast and sensitive methodology has been developed for the evaluation of the 3,4-methylenedioxypyrovalerone (MDPV) consumed. Based on ion mobility spectrometry (IMS), MDPV was directly determined in nasal fluids with a limit of detection (LOD) in the order of 22 ng mL, which corresponds to an absolute amount of 33 ng of MDPV per swab. MDPV was also determined after liquid-liquid microextraction (LLME) in oral fluids to avoid matrix effects, obtaining a LOD value of 4.4 ng mL in oral fluid samples. The IMS spectrum for MDPV exhibited a peak with K = 1.210 ± 0.005 cmV s at a drift time of 14.62 ms, the total analysis time being 4.5 min per oral fluid and 1.5 min per nasal fluid sample. Samples must be analyzed within 24 h following collection and dissolution in 2-propanol, based on the complementary stability studies. [Figure not available: see fulltext.] [ABSTRACT FROM AUTHOR]