Back to Search Start Over

A portable paper-based microfluidic platform for multiplexed electrochemical detection of human immunodeficiency virus and hepatitis C virus antibodies in serum.

Authors :
Chen Zhao
Xinyu Liu
Source :
Biomicrofluidics. 2016, Vol. 10 Issue 2, p1-10. 10p.
Publication Year :
2016

Abstract

This paper presents a portable paper-based microfluidic platform for multiplexed electrochemical detection of antibody markers of human immunodeficiency virus (HIV) and hepatitis C virus (HCV) in serum samples. To our best knowledge, this is the first paper-based electrochemical immunosensing platform, with multiplexing and telemedicine capabilities, for diagnosing HIV/HCV co-infection. The platform consists of an electrochemical microfluidic paper-based immunosensor array (E-μPIA) and a handheld multi-channel potentiostat, and is capable of performing enzyme-linked immunosorbent assays simultaneously on eight samples within 20 min (using a prepared E-μPIA). The multiplexing feature of the platform allows it to produce multiple measurement data for HIV and HCV markers from a single run, and its wireless communication module can transmit the results to a remote site for telemedicine. The unique integration of paper-based microfluidics and mobile instrumentation renders our platform portable, low-cost, user-friendly, and high-throughput. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
19321058
Volume :
10
Issue :
2
Database :
Academic Search Index
Journal :
Biomicrofluidics
Publication Type :
Academic Journal
Accession number :
115096502
Full Text :
https://doi.org/10.1063/1.4945311