Back to Search
Start Over
Comparison of the performance of forward fill/flush and reverse fill/flush flow modulation in comprehensive two-dimensional gas chromatography.
- Source :
-
Journal of Chromatography A . Sep2016, Vol. 1466, p113-128. 16p. - Publication Year :
- 2016
-
Abstract
- The performances of forward flow fill and flush (FFF) and of reverse flow fill and flush (RFF) in flow modulated comprehensive two-dimensional gas chromatography (GC × GC) using the same volume of the sampling channel have been studied and compared. Sample models include a reference mixture of hydrocarbons at low concentration, a petroleum reformate product and the essential oil of Rosa damascena Miller. The latter samples contain solutes in different concentrations but some up to 30% allowing to study overloading phenomena in detail. For solutes injected at low quantity, the performance of FFF and RFF is similar. For solutes present in a sample at high quantity, RFF guarantees less broadening and spreading resulting in better quantitation. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 00219673
- Volume :
- 1466
- Database :
- Academic Search Index
- Journal :
- Journal of Chromatography A
- Publication Type :
- Academic Journal
- Accession number :
- 118181153
- Full Text :
- https://doi.org/10.1016/j.chroma.2016.08.032