Radiation dose reduction using 100-kVp and a sinogram-affirmed iterative reconstruction algorithm in adolescent head CT: Impact on grey-white matter contrast and image noise.

Authors :
Nagayama, Yasunori
Nakaura, Takeshi
Tsuji, Akinori
Urata, Joji
Furusawa, Mitsuhiro
Yuki, Hideaki
Hirarta, Kenichiro
Kidoh, Masafumi
Oda, Seitaro
Utsunomiya, Daisuke
Yamashita, Yasuyuki
Source :
European Radiology. Jul2017, Vol. 27 Issue 7, p2717-2725. 9p. 3 Black and White Photographs, 4 Charts.
Publication Year :
2017

Abstract

<bold>Objectives: </bold>To retrospectively evaluate the image quality and radiation dose of 100-kVp scans with sinogram-affirmed iterative reconstruction (IR) for unenhanced head CT in adolescents.<bold>Methods: </bold>Sixty-nine patients aged 12-17 years underwent head CT under 120- (n = 34) or 100-kVp (n = 35) protocols. The 120-kVp images were reconstructed with filtered back-projection (FBP), 100-kVp images with FBP (100-kVp-F) and sinogram-affirmed IR (100-kVp-S). We compared the effective dose (ED), grey-white matter (GM-WM) contrast, image noise, and contrast-to-noise ratio (CNR) between protocols in supratentorial (ST) and posterior fossa (PS). We also assessed GM-WM contrast, image noise, sharpness, artifacts, and overall image quality on a four-point scale.<bold>Results: </bold>ED was 46% lower with 100- than 120-kVp (p < 0.001). GM-WM contrast was higher, and image noise was lower, on 100-kVp-S than 120-kVp at ST (p < 0.001). CNR of 100-kVp-S was higher than of 120-kVp (p < 0.001). GM-WM contrast of 100-kVp-S was subjectively rated as better than of 120-kVp (p < 0.001). There were no significant differences in the other criteria between 100-kVp-S and 120-kVp (p = 0.072-0.966).<bold>Conclusions: </bold>The 100-kVp with sinogram-affirmed IR facilitated dramatic radiation reduction and better GM-WM contrast without increasing image noise in adolescent head CT.<bold>Key Points: </bold>• 100-kVp head CT provides 46% radiation dose reduction compared with 120-kVp. • 100-kVp scanning improves subjective and objective GM-WM contrast. • Sinogram-affirmed IR decreases head CT image noise, especially in supratentorial region. • 100-kVp protocol with sinogram-affirmed IR is suited for adolescent head CT. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
09387994
Volume :
27
Issue :
7
Database :
Academic Search Index
Journal :
European Radiology
Publication Type :
Academic Journal
Accession number :
123476725
Full Text :
https://doi.org/10.1007/s00330-016-4679-6