Measurement of Heating Rates in a Microscopic Surface-Electrode Ion Trap.

Authors :
Jiu-Zhou He
Lei-Lei Yan
Liang Chen
Ji Li
Mang Feng
Source :
Chinese Physics Letters. Jul2017, Vol. 34 Issue 6, p1-1. 1p.
Publication Year :
2017

Abstract

We report measurement of heating rates ofions confined in our home-made microscopic surface-electrode trap by a Doppler recooling method. The ions are trapped with approximatelyabove the surface, and are subjected to heating due to various noises in the trap. There are 3–5 ions involved to measure the heating rates precisely and efficiently. We show the heating rates in variance with the number and the position of the ions as well as the radio-frequency power, which are helpful for understanding the trap imperfection. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
0256307X
Volume :
34
Issue :
6
Database :
Academic Search Index
Journal :
Chinese Physics Letters
Publication Type :
Academic Journal
Accession number :
123590187
Full Text :
https://doi.org/10.1088/0256-307X/34/6/063701