Use of a heated graphite scrubber as a means of reducing interferences in UV-absorbance measurements of atmospheric ozone.

Authors :
Turnipseed, Andrew A.
Andersen, Peter C.
Williford, Craig J.
Ennis, Christine A.
Birks, John W.
Source :
Atmospheric Measurement Techniques. 2017, Vol. 10 Issue 6, p2253-2269. 17p.
Publication Year :
2017

Abstract

A new solid-phase scrubber for use in conventional ozone (O3) photometers was investigated as a means of reducing interferences from other UV-absorbing species and water vapor. It was found that when heated to 100-130 °C, a tubular graphite scrubber efficiently removed up to 500 ppb ozone and ozone monitors using the heated graphite scrubber were found to be less susceptible to interferences from water vapor, mercury vapor, and aromatic volatile organic compounds (VOCs) compared to conventional metal oxide scrubbers. Ambient measurements from a graphite scrubberequipped photometer and a co-located Federal equivalent method (FEM) ozone analyzer showed excellent agreement over 38 days of measurements and indicated no loss in the scrubber's ability to remove ozone when operated at 130 °C. The use of a heated graphite scrubber was found to reduce the interference from mercury vapor to ≤3% of that obtained using a packed-bed Hopcalite scrubber. For a series of substituted aromatic compounds (ranging in volatility and absorption cross section at 253.7 nm), the graphite scrubber was observed to consistently exhibit reduced levels of interference, typically by factors of 2.5 to 20 less than with Hopcalite. Conventional solid-phase scrubbers also exhibited complex VOC adsorption and desorption characteristics that were dependent upon the relative humidity (RH), volatility of the VOC, and the available surface area of the scrubber. This complex behavior involving humidity is avoided by use of a heated graphite scrubber. These results suggest that heated graphite scrubbers could be substituted in most ozone photometers as a means of reducing interferences from other UV-absorbing species found in the atmosphere. This could be particularly important in ozone monitoring for compliance with the United States (U.S.) Clean Air Act or for use in VOC-rich environments such as in smog chambers and monitoring indoor air quality. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
18671381
Volume :
10
Issue :
6
Database :
Academic Search Index
Journal :
Atmospheric Measurement Techniques
Publication Type :
Academic Journal
Accession number :
123801771
Full Text :
https://doi.org/10.5194/amt-10-2253-2017