Measuring fluorescence polarization with a dichrometer.

Authors :
Sutherland, John C.
Source :
Analytical Biochemistry. Sep2017, Vol. 532, p9-11. 3p.
Publication Year :
2017

Abstract

A method for obtaining fluorescence polarization data from an instrument designed to measure circular and linear dichroism is compared with a previously reported approach. The new method places a polarizer between the sample and a detector mounted perpendicular to the direction of the incident beam and results in determination of the fluorescence polarization ratio, whereas the previous method does not use a polarizer and yields the fluorescence anisotropy. A similar analysis with the detector located axially with the excitation beam demonstrates that there is no frequency modulated signal due to fluorescence polarization in the absence of a polarizer. [ABSTRACT FROM AUTHOR]